SM-FTN leaks at injection pod port 3 - WKB122509
Article number: 122509
SYMPTOMS
- Leaking at injection pod port 3
- Low or no peak areas
ENVIRONMENT
- ACQUITY UPLC H-Class Series
- Sample Manager - Flow Through Needle (SM-FTN)
CAUSE
Tubing has come out of the PEEK ferrule.
FIX or WORKAROUND
Push the PEEK tubing a centimeter through the PEEK ferrule and reseat it into the injection pod.
ADDITIONAL INFORMATION