SM-FTN leaks at injection pod port 3 - WKB122509

Article number: 122509

SYMPTOMS

  • Leaking at injection pod port 3
  • Low or no peak areas

ENVIRONMENT

  • ACQUITY UPLC H-Class Series
  • Sample Manager - Flow Through Needle (SM-FTN)

CAUSE

Tubing has come out of the PEEK ferrule.

FIX or WORKAROUND

Push the PEEK tubing a centimeter through the PEEK ferrule and reseat it into the injection pod.

ADDITIONAL INFORMATION

id122509, UPARHSMFTN, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO