ACQUITY UPLC Sample Manager - High carryover with Flow Through Needle PLUS (SM-FTN PLUS) - WKB121820
Article number: 121820
SYMPTOMS
- Carryover occurs only in SM-FTN with a sample for which no carryover is seen with other models.
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Using with Load Ahead mode On.
FIX or WORKAROUND
Using with Load Ahead mode Off.
ADDITIONAL INFORMATION
Is there documentation that explains the "load ahead" option with the ACQUITY UPLC Sample Manager? - WKB8852