ACQUITY UPLC Sample Manager - High carryover with Flow Through Needle PLUS (SM-FTN PLUS) - WKB121820

Article number: 121820

SYMPTOMS

  • Carryover occurs only in SM-FTN with a sample for which no carryover is seen with other models.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Using with Load Ahead mode On.

FIX or WORKAROUND

Using with Load Ahead mode Off.