ACQUITY UPLC Sample Manager - High carryover with Flow Through Needle PLUS (SM-FTN PLUS) - WKB121820

Article number: 121820

SYMPTOMS

  • Carryover occurs only in SM-FTN with a sample for which no carryover is seen with other models.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Using with Load Ahead mode On.

FIX or WORKAROUND

Using with Load Ahead mode Off.

FTN.png

ADDITIONAL INFORMATION

Is there documentation that explains the "load ahead" option with the ACQUITY UPLC Sample Manager? - WKB8852

id121820, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO