ACQUITY UPLC Sample Manager - High carryover with Flow Through Needle PLUS (SM-FTN PLUS) - WKB121820
Article number: 121820
SYMPTOMS
- Carryover occurs only in SM-FTN with a sample for which no carryover is seen with other models.
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Using with Load Ahead mode On.
FIX or WORKAROUND
Using with Load Ahead mode Off.