Contaminant ions observed in LC-MS runs that use preslit PTFE or silicone septa vial caps - WKB12032
Article number: 12032
SYMPTOMS
- Contaminant ions observed in LC-MS runs that use preslit PTFE or silicone septa vial caps
- Contaminant ions are observed when doing replicate injections; the first injection is clean
- Contaminant ions are generally siloxanes
ENVIRONMENT
- LC-MS with Waters vial caps (186005666CV or 186000327C)
CAUSE
After the first injection, the solvent on the outside of the needle reacts with the outside (top) and slit of the septum.
FIX or WORKAROUND
The following are better vial caps to use:
186004169 - Polyethylene septumless vial cap
186007198 - PTFE septum in cap
186005826 - PTFE silicone (the lowest bleed PTFE Sil septum Waters supplies)
ADDITIONAL INFORMATION
When using 186004169 or 186007198, the hole in the cap will stay open after piercing, so samples may evaporate over time.
