Air bubble in 2695 sample syringe - WKB120305

Article number: 120305

SYMPTOMS

  • Air bubble in sample syringe
  • Replacing and purging the syringe do not resolve the issue

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Air is trapped in the flow path.

FIX or WORKAROUND

Purge the syringe from the diagnostics screen as follows:

  1. On the Alliance screen, navigate to Diagnostics > Others > Motors and Valves.
  2. Change Valves 1 and 3 to open position.
  3. Change the needle wash position to stream.
  4. Repeatedly move syringe to "Home" and "Empty" positions until the bubble dislodges.
  5. Exit the Diagnostics page.

ADDITIONAL INFORMATION

 