Air bubble in 2695 sample syringe - WKB120305
Article number: 120305
SYMPTOMS
- Air bubble in sample syringe
- Replacing and purging the syringe do not resolve the issue
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Air is trapped in the flow path.
FIX or WORKAROUND
Purge the syringe from the diagnostics screen as follows:
- On the Alliance screen, navigate to Diagnostics > Others > Motors and Valves.
- Change Valves 1 and 3 to open position.
- Change the needle wash position to stream.
- Repeatedly move syringe to "Home" and "Empty" positions until the bubble dislodges.
- Exit the Diagnostics page.