Skip to main content
Waters

"Peak location not near 397 nm" error message with 2475 FLR detector - WKB119714

Article number: 119714

SYMPTOMS

  • Error message: "Peak location not near 397 nm" 
  • Cleaned the flow cell with methanol, then flushed the flow cell with water for additional 10 minutes; error remains and wavelength is only 390 nm 

ENVIRONMENT

  • 2475 FLR Detector

CAUSE

Faulty flow cell

FIX or WORKAROUND
FLR.jpg

  1. Unscrew the in- and outlet tubing (green mark)
  2. Release the 3 screw (red marks) - they cannot be removed completely. 
  3. Replace the flow cell (205001034). 
  4. Flush the new flow cell with HPLC-grade water. 
  5. Perform normalize unit on the detector.

ADDITIONAL INFORMATION

 

id119714, 2475, 2475B, A-10FL, A-30FL, UPFLRARC, UPFLRDET, UPPFLR

Not able to find a solution? Click here to request help.