"Peak location not near 397 nm" error message with 2475 FLR detector - WKB119714
Article number: 119714
SYMPTOMS
- Error message: "Peak location not near 397 nm"
- Cleaned the flow cell with methanol, then flushed the flow cell with water for additional 10 minutes; error remains and wavelength is only 390 nm
ENVIRONMENT
- 2475 FLR Detector
CAUSE
Faulty flow cell
FIX or WORKAROUND

- Unscrew the in- and outlet tubing (green mark)
- Release the 3 screw (red marks) - they cannot be removed completely.
- Replace the flow cell (205001034).
- Flush the new flow cell with HPLC-grade water.
- Perform normalize unit on the detector.
ADDITIONAL INFORMATION
id119714, 2475, 2475B, A-10FL, A-30FL, UPFLRARC, UPFLRDET, UPPFLR