Error "Shutter failed to home" on ACQUITY UPLC PDA Detector - WKB1193

Article number: 1193

SYMPTOMS

  • The error "PDA Detector: Shutter failed to home" occurs
  • Recycling the power does not correct it
  • Shutter assembly could not home when 2998 is powered-on after a shutdown

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • 2998 Photodiode Array Detector
  • Arc Premier PDA

CAUSE

Air bubbles in the flow cell

FIX or WORKAROUND

  1. Power off the Detector
  2. Close the ACQUITY Console
  3. Flush the flow cell for 5-10 minutes with Organic solvent (100% Methanol or 100% Acetonitrile)
  4. Power on the Detector while solvent is flowing though the flow cell.

ADDITIONAL INFORMATION