How to acquire data in 410 mode (with 2414 RID and Empower 3) - WKB119360
Article number: 119360
OBJECTIVE or GOAL
Perform data acquisition in 410 mode.
ENVIRONMENT
- 2414 Refractive Index Detector
- Empower 3
PROCEDURE
- Select 410 mode on the 2414 detector body panel. Press Shift → DIAG on the 2414 detector body panel, select 410 for Detector Mode, and press the HOME key.
- Create a new chromatographic system including W410. (For procedures, see How to create a new chromatographic system in Empower 3 - WKB51287.)
- From the list of available instruments, select W410 and the other instruments you are using in combination with the 2414 detector.
- Reboot instrument and LACE
- Select the chromatographic system created in step 2 to open the Run Sample window, and create a new instrument method.
- Confirm that the instrument icon is displayed as W410.