Low sensitivity when using APGC - WKB118664

Article number: 118664

SYMPTOMS

  • Low signal on SST samples observed in MRMs    

ENVIRONMENT

  • APGC

CAUSE

GC inlet maintenance required

FIX or WORKAROUND

Perform GC inlet maintenance (change gold seal, septum, and liner).

ADDITIONAL INFORMATION

 

id118664, APGC