Check Valve failure - WKB117811

Article number: 117811

SYMPTOMS

  • Lost Prime error
  • Pressure near zero during operation
  • Pressure Fluctuation

ENVIRONMENT

  • Alliance HPLC System
  • Arc HPLC
  • ACQUITY Arc
  • ACQUITY UPLC
  • 2525/2545 Binary Gradient Module
  • 2535/2545/2555 Quaternary Gradient Module
  • ACQUITY UPLC H-Class PLUS QSM

CAUSE

Primary Check Valve blockage or faulty

FIX or WORKAROUND

  1. Remove the check valve(s).
  2. Sonicate the check valve(s) in water, then in methanol, and then in water again.
  3. Replace check valve(s) if necessary

ADDITIONAL INFORMATION

The sapphire or ceramic ball inside the check valve can get stuck. Hitting it on a bench top will often free it. Avoid hitting when the valve is connected to the pump head or check valve housings. 

Not able to find a solution? Click here to request help.