ACQUITY and MS communication failures on all modules - WKB117590

Article number: 117590

SYMPTOMS

  • Communication failure occurs on all modules randomly or constantly
  •  NetEvaluator fails for all modules
  • MS may show no comms or contunually dropping comms
  • Troubleshoot by connecting each module directly to the PC or switch box
  • Inlet logs show all units failing with sample manager failing first

ENVIRONMENT

  • ACQUITY UPLC Sample Manager 
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY APC Isocratic Solvent Manager
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY Sample Manager FL
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • Mass spec if present

CAUSE

Faulty Ethernet cable to Sample manager or to switch box 

FIX or WORKAROUND

Replace Ethernet Cable - Regular (441000372).

ADDITIONAL INFORMATION