1525 Binary HPLC Pump pressure is 0 - WKB117439
Article number: 117439
SYMPTOMS
- Pressure reading is 0 when pump is running
- loss of prime
ENVIRONMENT
- 1515 Isocratic HPLC Pump
- 1525 Binary HPLC Pump
- 2525/2545 Binary Gradient Module
- 2535/2545/2555 Quaternary Gradient Module
CAUSE
Pump is not primed, and air bubbles are in the check valves.
FIX or WORKAROUND
- Turn the vent valve to the right to waste.
- Set the flow rate to 5 mL/min.
- Use a syringe to pull solvent through the prime valve.
- Ensure that solvent flows consistently out of the waste line.
ADDITIONAL INFORMATION
id117439, 1515, 1525, 1525U, 1525UIVD, checkvalve