Lower biochem valve leaks on 2767 - WKB115721

Article number: 115721

SYMPTOMS

  • When priming the 2767, a leak is seen coming from the hole at the center of the lower biochem valve

ENVIRONMENT

  • 2767

CAUSE

Blocked needle

FIX or WORKAROUND

Replace the vented probe (700001387).

ADDITIONAL INFORMATION