With I-Class Sample Manager-FTN, “Calibration pin does not exist” error appears when performing calibrate rotary tray - WKB115353
Article number: 115353
SYMPTOMS
- Error message "Rotary Tray not calibrated"
- Error “Calibration pin does not exist” when performing calibrate rotary tray
ENVIRONMENT
-
ACQUITY I-Class Sample Manager-FTN
CAUSE
Faulty Datum Base with Theta Drive (700009452)
FIX or WORKAROUND
-
Replace the Datum Base with Theta Drive (700009452).
-
Recalibrate the rotary tray and R axis.
ADDITIONAL INFORMATION
The needle z axis may also need recalibration if you see this error.
