How do I use Helium sparge with preparative chromatography? - WKB11501
Article number: 11501
OBJECTIVE or GOAL
To show an approach to setting up Helium sparge for inline mixing of two mobile phases.
ENVIRONMENT
2535/2545/2555 Quaternary Gradient Module
PROCEDURE
- Obtain and secure a tank of 0.9990 purity Helium with a 2-stage regulator.
- Connect a stainless steel line with 0.040 ID from the regulator to the bulkhead fitting on the pump.
- Use three hole caps (WAT062479) for 1 L bottles or (WAT062341) for 4 L bottles.
- Place the solvent and sparge line through the cap and into the mobile phase bottle.
- Place a solvent filter upside-down on top of the cap through the third hole.
- Set second stage of the regulator to 100 psi.
- Turn the sparge on to 100% flow for 20 minutes for every 1 L of mobile phase.
- With desired mobile phase flow set, adjust Helium flow to 20%.
ADDITIONAL INFORMATION
You can use this procedure for all pumps without an inline degasser.