ACQUITY module missing from ACQUITY Console - WKB114433
Article number: 114433
SYMPTOMS
- No communications with an ACQUITY module
- After clearing the missing module from Waters DHCP and rebooting, the module does not repopulate DHCP
ENVIRONMENT
- MassLynx
- ACQUITY APC
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY APC Sample Manager - FTN
- ACQUITY Arc
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC Column Manager
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- PATROL UPLC Process Analysis System
- ACQUITY UPLC Process Sample Manager
- ACQUITY UPLC Column Heater
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC Sample Organizer
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC FLR Detector
- ACQUITY UPLC RI Detector
- ACQUITY UPLC Online SPE Man
CAUSE
Loose Ethernet cable
FIX or WORKAROUND
Reseat the Ethernet cable.
ADDITIONAL INFORMATION
id114433, comms, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM