Intermittently poor reproducibility on an ACQUITY H-Class - WKB113694

Article number: 113694

SYMPTOMS

  • Intermittently poor reproducibility
  • Issue remains after maintenance 
  • Dips in pressure trace observed
  • Passes decay test

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio

CAUSE

Faulty waste outlet check valve

FIX or WORKAROUND

Replace the outlet check valve (410003180; WASTE VALVE, INLINE, 1/4-28 THREADS).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.