ACQUITY UPLC system over pressure - WKB113364
SYMPTOMS
- The system pressure is over the MAX limit with flow rate established in analytical method
ENVIRONMENT
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class
- ACQUITY UPLC
CAUSE
Clog in system flow path
FIX or WORKAROUND
- Find the clogged tubings and replace them
    - 
        Remove the probe tubing from the MS divert valve. 
- 
        Remove the LC tubing from the MSdivert valve. 
- 
        Remove the other end of the LC tubing from the UV detector output. 
- 
        Remove the tubing from the UV detector input. 
- 
        Remove the other end of the UV detector input tubing from the column or column manager. 
- 
        If a column manager is present, switch it to bypass. 
- 
        If the issue is present only when not in bypass, or if not column manager prsent, remove the column. 
- 
        If a column manager is present, remove the active preheater from the inlet valve of the manager. 
- 
        Remove the active preheater or column manager input tubing from the injection valve. 
- 
        Switch the injection valve to both the load and inject positions. 
- 
        Remove the pump outlet tubing from the inject valve. 
- 
        Remove the pump outlet tubing from the mixer. 
- 
        Remove the mixer. 
- 
        Remove the mixer inlet tubing from the vent valve output. 
- 
        Remove the pump head outlet tubing from the vent valve inlet. 
- 
        Remove the pump head outlet tubing from the accumlator check valve. By this point, you will either have found the blockage or isolated it to the pump head 
 
- 
        
- Flush the system with Magic Mix (25% water, 25% methanol, 25% acetonitrile, 25% isopropanol).
- Replace the column.
ADDITIONAL INFORMATION
Bacteria will grow in pure aqueous mobile phase if the system is shut down for more than a week.
id113364, APH, checkvalve, eluent, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO
 
 