Skip to main content
Waters

ACQUITY UPLC H-Class does not inject the sample when going into the vial - WKB112664

Article number: 112664

SYMPTOMS

  • ACQUITY UPLC H-Class does not inject the sample when going into the vial
  • Syringe leak test fails

ENVIRONMENT

  • ACQUITY UPLC H-Class Flow Through Needle (FTN)

CAUSE

Air in the sample syringe

FIX or WORKAROUND

Purge the sample syringe 25 times with 100% methanol.

ADDITIONAL INFORMATION

 

id112664, UPBINARY

Not able to find a solution? Click here to request help.