How to change the sample syringe draw rate on ACQUITY UPLC Sample Manager - WKB112505

Article number: 112505

OBJECTIVE or GOAL

Change the sample syringe draw rate to either slow down or speed up sample uptake during injections.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager 
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

PROCEDURE

  1. In the method editor, select the sample manager.
  2. Select the Advanced tab in the bottom-right of the window.
    1. clipboard_eb7d97cd4caaccb530b995aef8cffdba7.png
  3. The first line in the Advanced window will be sample syringe draw rate. Specify the value in µL/min instead of automatic.
    1. clipboard_e79c79a7fb02133ceee61f490fe797a83.png

ADDITIONAL INFORMATION

clipboard_e6654b81808b4b2f6dfc39e3fafe97a86.png

id112505, A-30SM, UPBSMIVD, UPIBSMIVD, UPISMFTN, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPPFTN, UPSMFTN, UPSMFTN+, UPSMIVD, UPSMPMGR, UPSMPMGR