Noisy baseline seen using THF on ACQUITY RI detector in an APC system - WKB112175
Article number: 112175
SYMPTOMS
- Noisy baseline observed using THF on ACQUITY RI detector in APC system
- Solvent manager dynamic leak test successful but with a higher than expected leak rate (200 nL/min) for the primary head
- Consistent negative spikes at regular intervals in RI trace at low flow
ENVIRONMENT
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY UPLC RI Detector
CAUSE
The primary check valve is not operating effectively at low pressures when pumping THF.
FIX or WORKAROUND
Replace the primary check valve (700004139).
ADDITIONAL INFORMATION
id112175, A-30RI, checkvalve, DLT, UPASMFTN, UPMASM, UPRI, UPRIARC