Noisy baseline seen using THF on ACQUITY RI detector in an APC system - WKB112175
Article number: 112175
SYMPTOMS
- Noisy baseline observed using THF on ACQUITY RI detector in APC system
- Solvent manager dynamic leak test successful but with a higher than expected leak rate (200 nL/min) for the primary head
- Consistent negative spikes at regular intervals in RI trace at low flow
ENVIRONMENT
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY UPLC RI Detector
CAUSE
The primary check valve is not operating effectively at low pressures when pumping THF.
FIX or WORKAROUND
Replace the primary check valve (700004139).