Sample Organizer error "Organizer Z Drive lost steps during plate transfer" - WKB111462
Article number: 111462
SYMPTOMS
- On setting sample list, Sample Organizer error "Organizer Z Drive lost steps during plate transfer" occurs
ENVIRONMENT
- ACQUITY UPLC Sample Organizer
CAUSE
The plate is not installed correctly on the sample organizer shelf or the lower sensor has become misaligned and is not seeing the shelves/plates properly
FIX or WORKAROUND
- Stop the batch.
- Power-off the organizer.
- Remove the plate.
- Check the positioning of the sensor, adjust if necessary
- Power-on the sample organizer.
- Reset the sample organizer / sample manager in the console.
- Install the plate.
- Select Scan and store the shelf layout from the sample organizer console drop-down menu (Configure > Scan and store shelf layout).
- Start the sample list and confirm plate transfer.
ADDITIONAL INFORMATION
