"Needle/seal move limit exceeded" error on ACQUITY UPLC Sample Manager - Flow Through Needle - WKB110079

Article number: 110079

SYMPTOMS

"Needle/seal move limit exceeded" error message on ACQUITY UPLC Sample Manager - Flow Through Needle.

ENVIRONMENT

ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

ACQUITY UPLC H-Class Sample Manager - Flow Through Needle

CAUSE

Damage to the needle

FIX or WORKAROUND

Replace needle (see GPL).

ADDITIONAL INFORMATION

 

id110079, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO