SYMPTOMS

  • Seal Pack Geometry Fault (2) error

ENVIRONMENT

  • 2695 Separations Module

CAUSE

Needle top sensor needs adjustment up or down.

FIX or WORKAROUND

Adjust the needle top sensor until the Seal Pack calibration works.

ADDITIONAL INFORMATION

See error code information at: