"Cannot set flow at this time" error message when trying to set the flow on a 1515 pump - WKB10926

Article number: 10926

SYMPTOMS

  • "Cannot set flow at this time" error message when trying to set the flow on a 1515 pump through Breeze 
  • All instruments OK?  Yes in the node properties instruments tab, except for the 1515
  • Cannot ping the IP address of the 1515; all other instruments ping and get a reply
  • Changing the port on the Ethernet switch does not resolve the issue
  • Unplugging the Ethernet switch and rebooting the instruments and the computer does not resolve the issue

ENVIRONMENT

  • Breeze 2
  • 1515 Isocratic HPLC Pump
  • 2998 PDA Detector
  • 2414 RI Detector
  • Wyatt light scattering detector 
  • Eight-port Ethernet switch 

CAUSE

Bad Ethernet cable

FIX or WORKAROUND

Replace the Ethernet cable.

ADDITIONAL INFORMATION

For ethernet cable part numbers, see article What kind of Ethernet cable is recommended to connect Ethernet instruments and what is the part number? - WKB3373

id10926, 1515, 2414, 2414B, 2998, 2998B, 2998SFC, A-10PDA, A-10RI, BREEZESW, SUPBZ