Article number: 108135

SYMPTOMS

  • Noisy baseline observed with ACQUITY PDA detector
  • Signal-to-noise ratio low

ENVIRONMENT

  • ACQUITY UPLC PDA Detector

CAUSE

The lamp energy is low.

FIX or WORKAROUND

Replace the lamp.

Part Number for Acquity PDA/TUV 2489/2990 Perf Maint Kit  201000281, with ID Chip  700005269

ADDITIONAL INFORMATION

 

id108135, 2489, 2489B, 2489SFC, A-10UV, A-30PDA, UPAPDA, UPARPPDA, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA