ACQUITY UPLC H-Class Sample Manager FTN leaking - WKB108103

Article number: 108103

SYMPTOMS

  • Solvent is leaking all over injector pod and syringe             

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager FTN

CAUSE

The fitting at port 6 on the injector pod is loose.   

FIX or WORKAROUND

Tighten the fitting at port 6.