Article number: 101861

SYMPTOMS

  • No needle wash comes out of yellow waste line
  • No kinks in green or yellow line
  • Replacing needle wash pump does not resolve

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

CAUSE

Faulty needle wash valve

FIX or WORKAROUND

Replace the needle wash valve

V4 Plantium and Blue before 2018(WAT078800) or Enhanced V4 Solenoid Valve Assembly Blue Alliance after 2018 (# 700011808)