CPSO X-shuttle move error - WKB101841
Article number: 101841
SYMPTOMS
- X-shuttle move error observed on CPSO
- Plate is not transferred from Sample Manager back into the Sample Organizer
- Sample Organizer tries to load another plate into the Sample Manager when original plate is still in the Sample Manager
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC Sample Organizer - CPSO
CAUSE
X-transfer mechanism is missing the bolt/peg, which is needed to return sample plate from the SM to the CPSO.
FIX or WORKAROUND
- Power-off the Sample Organizer and manually transfer plates from the Sample Manager to the Sample Organizer.
- Manually move the X-transfer mechanism into the Sample Manager to gain access to the X-transfer mechanism.
- Using a T10 driver, tighten the bolt/peg back onto the X-transfer mechanism.

4. Complete full transfer of all plates to verify that the Sample Organizer is functioning as it should.
ADDITIONAL INFORMATION