CPSO X shuttle move error - WKB101841
Article number: 101841
SYMPTOMS
- X shuttle move error error observed on CPSO
- Plate not being transferred from Sample Manager back into the Sample Organizer
- Sample Organizer tries to load another plate into the Sample Manger when original plate is still in the Sample Manager
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC Sample Organizer - CPSO
CAUSE
X transfer mechanism bolt/peg missing which is needed to return sample plate from the SM to the CPSO.
FIX or WORKAROUND
- Power off Sample Organizer and manually transfer plates from the Sample Manager to the Sample Organizer.
- Manually move the X transfer mechanism into the Sample Manager to gain access to the X transfer mechanism.
- Using a T10 screwdriver, tighten the bolt/peg back onto the X transfer mechanism.

4. Complete full transfer of all plates to verify Sample Organizer is functioning as it should.