SYMPTOMS

  • High pressure during trapping on nanoACQUITY

ENVIRONMENT

  • nanoACQUITY UPLC

CAUSE

Blocked plumbing between nano tee and trap valve

FIX or WORKAROUND

Replace the plumbing (700001571).