Is there a function on the 2695 Separations Module that automatically stops flow when there is a leak or a column temperature rise? - WKB101087
Article number: 101087
ENVIRONMENT
- 2695 Separations Module
ANSWER
Because there is no leak sensor, there is no function to stop flow for a leak.
If the pressure falls below a certain value (25 psi), the flow will stop due to an error. Leaking is not considerable because at 25 psi flow is almost impossible.
If you set a "Low Pressure Limit" on the "Flow" tab of the instrument method, it is possible to stop the flow with an error when the pressure falls below that set value.
Also, if you set an acceptable range for column temperature on the "Temperature" tab of the instrument method, an error will occur if it is exceeded, but the run will not stop.