Skip to main content
Waters

Moisture in Sample Manager and Sample Organizer - WKB100927

Article number: 100927

SYMPTOMS

  • Needle arm home h/w fault
  • Moisture in Sample Manager and Sample Organizer

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer
  • ACQUITY UPLC Sample Manager 
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Indirect cause - rapid increase in ambient temperature and humidity due to air conditioning failure

Direct cause - humid air on the optic sensor

IMG_2609.jpg

FIX or WORKAROUND

Dry the sensor and the inside of the sample manager.

ADDITIONAL INFORMATION

The optimum temperature range of the laboratory is 19 to 22 °C (66 to 72 °F). - Site Preparation Guide

Sample manager environmental specifications - Operator's Guide

To keep the sample compartment from freezing, open its door only when necessary. (Opening the door admits humid air into the sample compartment, which causes condensation and freezing.) - Operator's Guide

Not able to find a solution? Click here to request help.