Flat peaks seen after changing lamp on a 2424 ELS Detector - WKB10039
Article number: 10039
SYMPTOMS
- 2424 lamp recently changed
- Peaks now show very high LSUs
- Peaks appear to be off the chart/flat-topped
ENVIRONMENT
- 2424 Evaporative Light Scattering Detector (ELSD)
CAUSE
The gain is set too high for the new lamp.
FIX or WORKAROUND
- Run the Auto Gain feature on the detector.
ADDITIONAL INFORMATION
To run Auto Gain, press DIAG 3, and then press Enter. See additional notes in the operator's guide (page 95; 3-31).
2424_Evaporative_Light_Scattering_Detector_Operator's_Guide_71500121802rb.pdf
