MS Quan result set is stuck in acquiring - WKB310412

Article number: 310412

SYMPTOMS

ENVIRONMENT

  • waters_connect
  • MS Quan app (all versions up to and including 2.1.0)

CAUSE

One of the Levels has been assigned a value of zero (0). MS Quan does not support using standards with a concentration of zero.

FIX or WORKAROUND

  1. Change the value of the zero level standard to 0.00001.
  2. Reprocess the data.

ADDITIONAL INFORMATION

See also wkb310234.