What are anti-virus exceptions required for waters_connect/UNIFI - WKB262086
ENVIRONMENT
- waters_connect
- UNIFI
- anti-virus / endpoint protection software
ANSWER
Note that this is not an exhaustive list of all the requirements and can change depending on software version. Please refer to the "Site Preparation Guide" for the list specific to your version.
- For real-time antivirus scanning, exclude all waters_connect-related folders and their sub-directories noted below.
- Real-time antivirus scanners mistake normal software functionality for virus activity, causing data buffering and or timeout problems.
- Ensure that scheduled antivirus scans do not occur during any data acquisition processes.
- This includes any advanced heuristics antivirus scanning engines such as Crowdstrike, Sentinel One, and similar products.
- Note: Learning AI features may also need to be disabled as Waters have many customers where this learned to block or deep scan our executables despite the exceptions.
Server
C:\ProgramData\Waters (hidden)
C:\Program Files\Waters
C:\Program Files\Waters Message Bus
C:\Program Files(x86)\Waters
<APP DRIVE>:\Waters
<APP DRIVE>:\Program Files\Waters
<APP DRIVE>:\Program Files(x86)\Waters
<BACKUP DRIVE>:\FRA
LND
C:\ProgramData\Waters (hidden)
<INSTALL DRIVE>:\Program Files\Waters
<INSTALL DRIVE>:\Program Files(x86)\Waters
Client
C:\ProgramData\Waters (hidden)
<INSTALL DRIVE>:\Program Files\Waters
*******************************************************************
Workstation
C:\ProgramData\Waters (hidden)
C:\Program Files\Waters
C:\Program Files\Waters Message Bus
C:\Program Files(x86)\Waters
<DATABASE DRIVE>:\Waters (second largest drive)
<BACKUP DRIVE>:\FRA (largest drive)
ADDITIONAL INFORMATION