"Hosted Content Error" when opening or creating a method - WKB205030
Article number: 205030
SYMPTOMS
- "Hosted Content Error" when opening or creating a method
- "WorkAreaItem/Waters.AnalysisMethod.UI:OutputsReportingItem"
- "/Waters.AnalysisMethod.UI:OutputsReportingItem"

ENVIRONMENT
- waters_connect with UNIFI 1.9.13
- waters_connect with UNIFI 2.1.2
- waters_connect with UNIFI 3.0.0
- waters_connect 3.1.0
CAUSE
The Printer Spooler Service is disabled.
INFRED-281
FIX or WORKAROUND
- Re-enable and start the Printer Spooler Service.
- Ensure that there are no local IT policies applied that will disable Printer Spooler Service
ADDITIONAL INFORMATION
- You can confirm the service is truly running by running netstat from command line on the server
c:\>netstat -anob
- This should report the spoolersv.exe is running