Skip to main content
Waters

Which Agilent modules does HPLC CONNECT support? - WKB319104

Article number: 319104

ENVIRONMENT

  • HPLC CONNECT
  • ASTRA 
  • HPLC DASHBOARD application
  • Agilent HPLC pump
  • Agilent autosampler
  • Agilent UV/Vis detectors (MWD, DAD, VWD)
  • Agilent RI detector (limited control via HPLC DASHBOARD)
  • Agilent fraction collectors (ASTRA 8.2+ with HPLC CONNECT 3)

ANSWER

HPLC CONNECT supports control of Agilent HPLC pump, autosampler, and UV/Vis detectors (MWD, DAD, VWD). With HPLC CONNECT 3 and ASTRA 8.2 or higher, fraction collector modules are also supported. The HPLC DASHBOARD application can control additional Agilent modules listed on the supported hardware list. ASTRA cannot control Agilent RI detectors during a run, but it can fully control Optilab RI detectors. Fraction collector parameter changes during a run are not currently supported by ASTRA.

ADDITIONAL INFORMATION

HPLC CONNECT allows ASTRA to control the HPLC pump, autosampler, and UV/Vis (MWD, DAD, or VWD) modules. HPLC CONNECT 3 adds fraction collector module support. See HPLC CONNECT Supported Hardware.pdf for a list of supported Agilent modules and information about digital UV support.

The HPLC DASHBOARD (a standalone application) can control other Agilent modules as long as they are on the list of supported modules. ASTRA will not be able to control the Agilent RI and cannot, for example, purge at the end of a run. ASTRA has full control over an Optilab RI detector.

Fraction collector support is available onlyfor ASTRA 8.2 and higher with HPLC CONNECT 3. For HPLC CONNECT 2, there are several models of fraction collector listed as supported, and if a specific collector is on the list, HPLC DASHBOARD might have access to this extended functionality via the Actions “Control…” Depending on what parameters are necessary for a run, it might be possible to set initial parameters at run start. However, ASTRA would not be able to perform fraction collector parameter changes while the run is in progress. This is a current feature request.

For the other types of Agilent detectors on the supported module list, the control would be what the Agilent API makes available to HPLC DASHBOARD. The functions would be similar to what Instant Pilot allowed. Currently, ASTRA does not support modules other than the HPLC pump, autosampler, MWD/VWD/DAD, and, in ASTRA 8.2 and higher with HPLC CONNECT 3, fraction collectors.

Not able to find a solution? Click here to request help.