In UNIFI, a post-translational modification configured in the science library is not reflected in the processing results - WKB96419
Article number: 96419
SYMPTOMS
- Pyroglutamylation of the heavy chain N-terminal Q of the antibody is configured in the science library
- Post-translational modification (pyroglutamylation of N-terminal Q) is not reflected on the Review screen
ENVIRONMENT
- UNIFI 1.9.4
CAUSE
Post-translational modifications are configured in the science library.
FIX or WORKAROUND
- Do not configure post-translational modifications (other than disulfide) in the science library.
- Configure post-translational modifications such as pyroglutamylation of the N-terminal Q of the heavy chain to “Fixed” or “Variable” in Analysis Specific Settings > Amino Acid Modifications in the analysis method.
ADDITIONAL INFORMATION