UNIFI: How to display the background subtracted spectrum? - WKB90037

Article number: 90037

OBJECTIVE or GOAL

  • Display background-subtracted spectrum on the Investigate screen

ENVIRONMENT

  • UNIFI

PROCEDURE

  1. Open the Investigate screen.
  2. Display the target chromatogram.
  3. Choose “Process Options”.
  4. Select Combine spectra from the pull-down menu.
  5. In “Times to average”, enter the range of retention time for which you want to integrate spectra.
  • It can be entered automatically by right-dragging on the Chromatogram screen.
  1. Check the “Times to subtract” and enter the range of the background retention time.
  • It can be entered automatically by right-dragging on the Chromatogram screen.
  • Make the time window of the entered “Times to average” and the “Times to subtract” to be the same.
  1. Click Start.

ADDITIONAL INFORMATION

 