Hosted Content Error message when opening an analysis method in UNIFI and Quantify/Calibrate settings are empty - WKB88440

Article number: 88440

SYMPTOMS

  • Error message when opening an analysis method:

  • Once analysis method is opened, there is no content in the "Quantify" and "Calibrate" sections of the Processing tab

ENVIRONMENT

  • UNIFI 1.9.4

CAUSE

There are duplicated and/or custom fields with same name in the custom field editor (see image below, for example).

FIX or WORKAROUND

  1. Remove the duplicated custom field, or rename the custom fields.

ADDITIONAL INFORMATION