How to extract and integrate the C12 and C13 peaks for a compound in UNIFI - WKB7395
Article number: 7395
OBJECTIVE or GOAL
Integrate the parent (C12) chromatographic peak and the first isotope (C13) chromatographic peak in UNIFI.
ENVIRONMENT
- UNIFI
PROCEDURE
- In the Analysis, go to the Investigate tab and select processing options in the top-right.
- Pull up the Chromatogram of the sample to be analyzed.
- Under Operations, select "Extract mass chromatogram" and input the mass range of the parent or C12 ion. Use a 10-ppm or smaller window and click the start button at the bottom. This extracts the chromatogram of the C12 ion.
- Select Integrate from the Operations drop-down menu.
- Right-click on the Integrated peak and Select properties.
- Go to Annotations and Peak Annotations in the Chromatogram Properties window.
- Expand Integrated LC System - Suitability Peak, scroll down to Area, and then select the blue down arrow to add it to Peak information.
- Select OK.
- The peak's area is now listed in the Chromatogram window for those peaks integrated.
- Repeat the extract mass chromatogram and integration for the C13 mass.