Cannot log into UNIFI - WKB6422
Article number: 6422
SYMPTOMS
- Typing the password in the UNIFI login screen does not open UNIFI
ENVIRONMENT
- UNIFI Workstation only
CAUSE
The OracleUNIFIservice is not running.
FIX or WORKAROUND
- Open Task Manager and click the Services tab
- Stop the Watservicehost and wait for the service to stop
- Right-click OracleUNIFIservice and click Start Service
- Right-click Watservicehost and click Start Service
- Wait for all 8 Watservicehost processes to start up
- Log in to UNIFI
ADDITIONAL INFORMATION
Set the service to automatically start by following these steps:
- In Windows Explorer, right-click Computer and select Manage.
- Expand the Services and Applications node.
- Click Services.
- Right-click OracleUNIFIservice and select Properties.
- In the Startup type list, select Automatic.
- Click Apply, and then click OK.