UNIFI filter not applied to report - WKB61601

Article number: 61601

SYMPTOMS

  • It is not possible to remove discovered peaks from the FLR chromatogram in the Unifi Report using a filter

  • It is possible to apply the same filter to a chromatogram in Review.

  • Tables can be filtered to display only identified peaks with mass confirmation in the report

ENVIRONMENT

  • UNIFI 1.9 SR3
  • UNIFI 1.9 SR4

CAUSE

The problems occurs because Unifi cannot apply filters in 'plot'  reporting objects. So the report can only show the full chromatogram, and not just a subset of filtered peaks

FIX or WORKAROUND

  1. None

ADDITIONAL INFORMATION

CRI342 and INFULA1464 have been created in relation to this issue.

 

id61601, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC