UNIFI backups are not working - WKB51649

Article number: 51649

SYMPTOMS

  • UNIFI full and incremental backups no longer run
  • UNIFI manual backups from WDM do not even start or create a backup set

ENVIRONMENT

  • UNIFI
  • UNIFI Workstation
  • Waters Database Manager

CAUSE

The oraclejobs password has expired.

FIX or WORKAROUND

  1. Go to the start button, right-click on Computer, and select Manage.
        IMG_2504.JPG
  2. Expand Local Users and Groups, click on Users, right-click on oraclejobs, and select Properties.
    IMG_2506.JPG
  3. Clear the "User must change password at next logon" option and add "User cannot change password" and "Password never expires".
    IMG_2507.JPGIMG_2508.JPG
  4. Click Apply, and then click OK.
  5. Perform a First_Full_LVL0 backup and ensure that it works.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.