Skip to main content
Waters

With UNIFI, ACQUITY FTN does not draw up sample before injecting - WKB50132

Article number: 50132

SYMPTOMS

  • No sample is drawn up from the vial during the injection cycle
  • Position of the needle looks to be off—it does not go to the top of the correct vial position
  • On starting an acquisition, system gets ready, injection process starts, and the needle moves to the correct vial position; the arm lowers to touch the vial and then bounces back up immediately without doing anything; it then moves back to needle seal and injection starts without drawing sample

ENVIRONMENT

  • UNIFI
  • ACQUITY - FTN

CAUSE

Unknown

FIX or WORKAROUND

  1. Open the analysis method and configure a different (incorrect) well plate.
  2. Save the analysis method and acquire a run.
  3. As expected, this results in an error.
  4. Reset the FTN.
  5. Configure the analysis method for the correct well plate.
  6. Save the analysis method and acquire.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.