With UNIFI, ACQUITY FTN does not draw up sample before injecting - WKB50132
Article number: 50132
SYMPTOMS
- No sample is drawn up from the vial during the injection cycle
- Position of the needle looks to be off—it does not go to the top of the correct vial position
- On starting an acquisition, system gets ready, injection process starts, and the needle moves to the correct vial position; the arm lowers to touch the vial and then bounces back up immediately without doing anything; it then moves back to needle seal and injection starts without drawing sample
ENVIRONMENT
- UNIFI
- ACQUITY - FTN
CAUSE
Unknown
FIX or WORKAROUND
- Open the analysis method and configure a different (incorrect) well plate.
- Save the analysis method and acquire a run.
- As expected, this results in an error.
- Reset the FTN.
- Configure the analysis method for the correct well plate.
- Save the analysis method and acquire.
ADDITIONAL INFORMATION