Start of data collection fails in UNIFI with ACQUITY UPLC I-Class PLUS FTN and BSM - WKB49100

Article number: 49100

SYMPTOMS

  • Mid sample set, the sample list stops and both the FTN and BSM show a "Failed to start collecting data" error message on the side panel

UNI failed to start collecting data.PNG

ENVIRONMENT

  • ACQUITY UPLC I-Class PLUS FTN
  • ACQUITY UPLC I-Class PLUS BSM
  • UNIFI

CAUSE

The active preheater (APH) drops out briefly and then connects again, resulting in a spike in the APH temperature.

FIX or WORKAROUND

Reseat or replace the active preheater (APH).

ADDITIONAL INFORMATION

 

id49100, RDA, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+