Elemental Composition fails, with an error in UNIFI - WKB47116

Article number: 47116

SYMPTOMS

  • Elemental Composition analysis fails, with an error in UNIFI

ENVIRONMENT

  • UNIFI 1.8.2

CAUSE

If there is an adduct with no chemical formula specified in the scientific library, Elemental Composition analysis fails.

FIX or WORKAROUND

Delete the faulty adduct entry in the scientific library.

  1. Navigate to "Scientific Library" > "Configure Scientific Library" > "Modifications" > "Adducts".
    clipboard_e6f75b00cb03a3fb06fc862a620e34a14.png
  2. Search for an entry that has no formula specified OR a formula that is written with small letters (e.g., +co2).
    clipboard_e602d345c323758e672b9028d41a21d45.png
  3. Delete the faulty entry.