Cannot delete folder in waters_connect - WKB261765

Article number: 261765

SYMPTOMS

  • Error message "Cannot delete folder" when trying to delete a folder in waters_connect
    clipboard_ef6667fce0b7066fcb539946c6bc27730.png

ENVIRONMENT

  • waters_connect 3.0.0 to 4.0.0

CAUSE

An item in the folder is locked for editing.

FIX or WORKAROUND

  1. Check all user accounts for work in progress and save and finish work in progress.
    Log in as an administrator, open UNIFI, go to the My Work tab, and select the all users button.
    clipboard_e995f1a1a9b92870249c9a871c6ca46cf.png
  2. Inspect user accounts for pending signatures.
  3. Open Explorer, select all items in a folder, right-click and select "Remove edit lock", and then delete the folder.
    clipboard_e7e1d8ba642b084e87539a7327123d404.png
  4. If the folder still cannot be deleted then contact GSS to remove remaining edit locks.
     

ADDITIONAL INFORMATION

This workflow does not work for sessions from web-based waters_connect applications such as MS-Quan or  LC-MS Toolkit.
Check wkb278482 if required.
 
If this does not work, contact GSS to remove the locks via sql edits.

Not able to find a solution? Click here to request help.