Cannot quantitate Tof MRM or MSMS data in UNIFI - WKB25291
Article number: 25291
SYMPTOMS
- No response or quantitation results after processing
- The component is clearly there
ENVIRONMENT
- UNIFI 1.7.X
- UNIFI 1.8.X
- UNIFI 1.9.2
CAUSE
The components in the analysis method were brought in via import from a library and have too much information.
FIX or WORKAROUND
In the analysis method Manage Components tab, keep data only in the following columns:
-
Expected RT
-
Calibration response factor
-
Extraction mode
-
Fragment charge
-
Fragment m/z
ADDITIONAL INFORMATION