Cannot quantitate Tof MRM or MSMS data in UNIFI - WKB25291

Article number: 25291

SYMPTOMS

  • No response or quantitation results after processing
  • The component is clearly there

ENVIRONMENT

  • UNIFI 1.7.X
  • UNIFI 1.8.X
  • UNIFI 1.9.2

CAUSE

The components in the analysis method were brought in via import from a library and have too much information.

FIX or WORKAROUND

In the analysis method Manage Components tab, keep data only in the following columns:

  • Expected RT

  • Calibration response factor

  • Extraction mode

  • Fragment charge

  • Fragment m/z

UNI MRM quan functions.PNG

ADDITIONAL INFORMATION

id25291, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC