No communication with Xevo G2-XS in UNIFI; error message "Instrument EPC failure" - WKB207222
Article number: 207222
SYMPTOMS
- No communication with Xevo G2-XS in UNIFI
- Error message "Instrument EPC failure"
- Instrument EPC is currently in an indeterminate status
ENVIRONMENT
- UNIFI Workstation
- UNIFI Network
- waters_connect
- Xevo G2-XS
- Xevo G2-XS QTof
CAUSE
Communication failure with EPC
FIX or WORKAROUND
- Reboot only the EPC, rather than a full electronic reboot.
- If EPC reboot does not resolve the issue, perform electronics reboot.
- Verify communication.
ADDITIONAL INFORMATION
The EPC can be rebooted using a piece of PEEK tubing to depress the small microswitch behind the sliding cover.
id207222, comms, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, XEVOG2XSTF