In UNIFI, cannot see more than 50 lines under Column History - WKB207206

Article number: 207206

SYMPTOMS

  • In the Acquisition Console, under the system's column selection, the Column History pane shows no more than 50 lines
  • Column History shows no filters or other toggles to view more than this number of entries

ENVIRONMENT

  • UNIFI 1.9.9
  • BioAccord System

CAUSE

This is a limitation of the column's eCord.

FIX or WORKAROUND

None. Only 50 sample sets can be stored on the eCord chip.

ADDITIONAL INFORMATION

Find more information in the corresponding column's Care and Use Manual in the Column Use Information section. An example is the ACQUITY UPLC and ACQUITY Premier BEH Columns Care and Use Manual.